Simulation guided design of the MRcollar: a MR compatible applicator for deep heating in the head and neck region

T. Drizdal, K. Sumser, GG. Bellizzi, O. Fiser, J. Vrba, GCV. Rhoon, DTB. Yeo, . Margarethus M Paulides

. 2021 ; 38 (1) : 382-392. [pub] -

Language English Country Great Britain

Document type Journal Article, Research Support, Non-U.S. Gov't

PURPOSE: To develop a head and neck hyperthermia phased array system compatible with a 1.5 T magnetic resonance (MR) scanner for noninvasive thermometry. METHODS: We designed a dielectric-parabolic-reflector antenna (DiPRA) based on a printed reflector backed dipole antenna and studied its predicted and measured performance in a flat configuration (30 mm thick water bolus and muscle equivalent layer). Thereafter, we designed a phased array applicator model ('MRcollar') consisting of 12 DiPRA modules placed on a radius of 180 mm. Theoretical heating performance of the MRcollar model was benchmarked against the current clinical applicator (HYPERcollar3D) using specific (3D) head and neck models of 28 treated patients. Lastly, we assessed the influence of the DiPRA modules on MR scanning quality. RESULTS: The predicted and measured reflection coefficients (S11) of the DiPRA module are below -20 dB. The maximum specific absorption rate (SAR) in the area under the antenna was 47% higher than for the antenna without encasing. Compared to the HYPERcollar3D, the MRcollar design incorporates 31% less demineralized water (-2.5 L), improves the predicted TC25 (target volume enclosed by 25% iso-SAR contour) by 4.1% and TC50 by 8.5%, while the target-to-hotspot quotient (THQ) is minimally affected (-1.6%). MR experiments showed that the DiPRA modules do not affect MR transmit/receive performance. CONCLUSION: Our results suggest that head and neck hyperthermia delivery quality with the MRcollar can be maintained, while facilitating simultaneous noninvasive MR thermometry for treatment monitoring and control.

References provided by Crossref.org

000      
00000naa a2200000 a 4500
001      
bmc21026250
003      
CZ-PrNML
005      
20211026133057.0
007      
ta
008      
211013s2021 xxk f 000 0|eng||
009      
AR
024    7_
$a 10.1080/02656736.2021.1892836 $2 doi
035    __
$a (PubMed)33682594
040    __
$a ABA008 $b cze $d ABA008 $e AACR2
041    0_
$a eng
044    __
$a xxk
100    1_
$a Drizdal, Tomas $u Hyperthermia Unit, Department of Radiotherapy, Erasmus MC Cancer Institute, Rotterdam, The Netherlands $u Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, Kladno, Czech Republic, Kladno, Czech Republic in Prague
245    10
$a Simulation guided design of the MRcollar: a MR compatible applicator for deep heating in the head and neck region / $c T. Drizdal, K. Sumser, GG. Bellizzi, O. Fiser, J. Vrba, GCV. Rhoon, DTB. Yeo, . Margarethus M Paulides
520    9_
$a PURPOSE: To develop a head and neck hyperthermia phased array system compatible with a 1.5 T magnetic resonance (MR) scanner for noninvasive thermometry. METHODS: We designed a dielectric-parabolic-reflector antenna (DiPRA) based on a printed reflector backed dipole antenna and studied its predicted and measured performance in a flat configuration (30 mm thick water bolus and muscle equivalent layer). Thereafter, we designed a phased array applicator model ('MRcollar') consisting of 12 DiPRA modules placed on a radius of 180 mm. Theoretical heating performance of the MRcollar model was benchmarked against the current clinical applicator (HYPERcollar3D) using specific (3D) head and neck models of 28 treated patients. Lastly, we assessed the influence of the DiPRA modules on MR scanning quality. RESULTS: The predicted and measured reflection coefficients (S11) of the DiPRA module are below -20 dB. The maximum specific absorption rate (SAR) in the area under the antenna was 47% higher than for the antenna without encasing. Compared to the HYPERcollar3D, the MRcollar design incorporates 31% less demineralized water (-2.5 L), improves the predicted TC25 (target volume enclosed by 25% iso-SAR contour) by 4.1% and TC50 by 8.5%, while the target-to-hotspot quotient (THQ) is minimally affected (-1.6%). MR experiments showed that the DiPRA modules do not affect MR transmit/receive performance. CONCLUSION: Our results suggest that head and neck hyperthermia delivery quality with the MRcollar can be maintained, while facilitating simultaneous noninvasive MR thermometry for treatment monitoring and control.
650    _2
$a hlava $x diagnostické zobrazování $7 D006257
650    12
$a vytápění $7 D006361
650    _2
$a lidé $7 D006801
650    12
$a indukovaná hypertermie $7 D006979
650    _2
$a magnetická rezonanční tomografie $7 D008279
650    _2
$a magnetická rezonanční spektroskopie $7 D009682
650    _2
$a krk $x diagnostické zobrazování $7 D009333
655    _2
$a časopisecké články $7 D016428
655    _2
$a práce podpořená grantem $7 D013485
700    1_
$a Sumser, Kemal $u Hyperthermia Unit, Department of Radiotherapy, Erasmus MC Cancer Institute, Rotterdam, The Netherlands
700    1_
$a Bellizzi, Gennaro G $u Hyperthermia Unit, Department of Radiotherapy, Erasmus MC Cancer Institute, Rotterdam, The Netherlands $u Department of Information Engineering, Infrastructures and Sustainable Energy, Universita Mediterranea di Reggio Calabria, Reggio di Calabria, Italy
700    1_
$a Fiser, Ondrej $u Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, Kladno, Czech Republic, Kladno, Czech Republic in Prague
700    1_
$a Vrba, Jan $u Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, Kladno, Czech Republic, Kladno, Czech Republic in Prague
700    1_
$a Rhoon, Gerard C van $u Hyperthermia Unit, Department of Radiotherapy, Erasmus MC Cancer Institute, Rotterdam, The Netherlands
700    1_
$a Yeo, Desmond T B $u Imaging and Bioelectronic Technologies, GE Global Research Centre, Niskayuna, NY, USA
700    1_
$a Margarethus M Paulides, $u Hyperthermia Unit, Department of Radiotherapy, Erasmus MC Cancer Institute, Rotterdam, The Netherlands $u Department of Electrical Engineering, Eindhoven University of Technology, Eindhoven, The Netherlands
773    0_
$w MED00002331 $t International journal of hyperthermia $x 1464-5157 $g Roč. 38, č. 1 (2021), s. 382-392
856    41
$u https://pubmed.ncbi.nlm.nih.gov/33682594 $y Pubmed
910    __
$a ABA008 $b sig $c sign $y p $z 0
990    __
$a 20211013 $b ABA008
991    __
$a 20211026133103 $b ABA008
999    __
$a ok $b bmc $g 1715076 $s 1146757
BAS    __
$a 3
BAS    __
$a PreBMC
BMC    __
$a 2021 $b 38 $c 1 $d 382-392 $e - $i 1464-5157 $m International journal of hyperthermia $n Int J Hyperthermia $x MED00002331
LZP    __
$a Pubmed-20211013

Find record

Citation metrics

Loading data ...

Archiving options

Loading data ...