Validation and inter-laboratory study of selective hydride generation for fast screening of inorganic arsenic in seafood
Jazyk angličtina Země Nizozemsko Médium print-electronic
Typ dokumentu časopisecké články
PubMed
30612652
DOI
10.1016/j.aca.2018.11.036
PII: S0003-2670(18)31392-8
Knihovny.cz E-zdroje
- Klíčová slova
- Certified reference materials, Inductively coupled plasma mass spectrometry, Inorganic arsenic, Interlaboratory comparison, Selective hydride generation, Speciation analysis,
- MeSH
- arsen analýza MeSH
- Decapoda (Crustacea) chemie MeSH
- Decapodiformes chemie MeSH
- hmotnostní spektrometrie metody MeSH
- kontaminace potravin analýza MeSH
- kyselina chlorovodíková chemie MeSH
- limita detekce MeSH
- měkkýši, korýši analýza MeSH
- mlži chemie MeSH
- mořské řasy chemie MeSH
- perchloráty chemie MeSH
- peroxid vodíku chemie MeSH
- potravní rybí proteiny analýza MeSH
- Squalidae MeSH
- vodík chemie MeSH
- zvířata MeSH
- Check Tag
- zvířata MeSH
- Publikační typ
- časopisecké články MeSH
- Názvy látek
- arsen MeSH
- kyselina chlorovodíková MeSH
- perchloráty MeSH
- Perchloric Acid MeSH Prohlížeč
- peroxid vodíku MeSH
- potravní rybí proteiny MeSH
- vodík MeSH
It is advisable to monitor and regulate inorganic arsenic (iAs) in food and feedstuff. This work describes an update and validation of a method of selective hydride generation (HG) with inductively coupled plasma mass spectrometry (ICP-MS) for high-throughput screening of iAs content in seafood samples after microwave-assisted extraction with diluted nitric acid and hydrogen peroxide. High concentration of HCl (8 M) for HG along with hydrogen peroxide in samples of a same concentration as used for extraction leads to a selective conversion of iAs to volatile arsine that is released and transported to the detector. A minor contribution from methylarsonate (≈20% to iAs) was found, while HG from dimethylarsinate, trimethylarsine oxide is substantially suppressed (less than 1% to iAs). Methodology was applied to Certified Reference Materials (CRMs) TORT-3, DORM-3, DORM-4, DOLT-4, DOLT-5, PRON-1, SQID-1 and ERM-CE278k, in some of them iAs has been determined for the first time, and to various seaweed samples from a local store. The results were always compared with a reference method and selectivity of iAs determination was evaluated. An inter-laboratory reproducibility was tested by comparative analyses of six fish and four seaweed samples in three European laboratories, with good agreement of the results. The method of HG-ICP-MS is sensitive (limit of detection 2 μg kg-1 iAs), well suited for screening of large number of samples and selective at iAs concentration levels at which maximum limits are expected to be set into EU legislation for marine samples.