Left bundle branch pacing compared to left ventricular septal myocardial pacing increases interventricular dyssynchrony but accelerates left ventricular lateral wall depolarization
Jazyk angličtina Země Spojené státy americké Médium print-electronic
Typ dokumentu srovnávací studie, časopisecké články, práce podpořená grantem
PubMed
33930549
DOI
10.1016/j.hrthm.2021.04.025
PII: S1547-5271(21)00402-1
Knihovny.cz E-zdroje
- Klíčová slova
- Depolarization duration, Dyssynchrony, His-bundle pacing, Left bundle branch pacing, Left ventricular septal myocardial pacing, Ultra-high-frequency electrocardiography,
- MeSH
- blokáda Tawarova raménka patofyziologie terapie MeSH
- elektrokardiografie metody MeSH
- funkce levé komory srdeční fyziologie MeSH
- Hisův svazek patofyziologie MeSH
- kardiostimulace umělá metody MeSH
- lidé MeSH
- mezikomorová přepážka patofyziologie MeSH
- následné studie MeSH
- prospektivní studie MeSH
- senioři MeSH
- srdeční komory patofyziologie MeSH
- Check Tag
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- práce podpořená grantem MeSH
- srovnávací studie MeSH
BACKGROUND: Nonselective His-bundle pacing (nsHBp), nonselective left bundle branch pacing (nsLBBp), and left ventricular septal myocardial pacing (LVSP) are recognized as physiological pacing techniques. OBJECTIVE: The purpose of this study was to compare differences in ventricular depolarization between these techniques using ultra-high-frequency electrocardiography (UHF-ECG). METHODS: In patients with bradycardia, nsHBp, nsLBBp (confirmed concomitant left bundle branch [LBB] and myocardial capture), and LVSP (pacing in left ventricular [LV] septal position without proven LBB capture) were performed. Timings of ventricular activations in precordial leads were displayed using UHF-ECG, and electrical dyssynchrony (e-DYS) was calculated as the difference between the first and last activation. Duration of local depolarization (Vd) was determined as width of the UHF-QRS complex at 50% of its amplitude. RESULTS: In 68 patients, data were collected during nsLBBp (35), LVSP (96), and nsHBp (55). nsLBBp resulted in larger e-DYS than did LVSP and nsHBp [- 24 ms (-28;-19) vs -12 ms (-16;-9) vs 10 ms (7;14), respectively; P <.001]. nsLBBp produced similar values of Vd in leads V5-V8 (36-43 ms vs 38-43 ms; P = NS in all leads) but longer Vd in leads V1-V4 (47-59 ms vs 41-44 ms; P <.05) as nsHBp. LVSP caused prolonged Vd in leads V1-V8 compared to nsHBp and longer Vd in leads V5-V8 compared to nsLBBp (44-51 ms vs 36-43 ms; P <.05) regardless of R-wave peak time in lead V5 or QRS morphology in lead V1 present during LVSP. CONCLUSION: nslbbp preserves physiological LV depolarization but increases interventricular electrical dyssynchrony. LV lateral wall depolarization during LVSP is prolonged, but interventricular synchrony is preserved.
Institute of Scientific Instruments the Czech Academy of Sciences Brno Czech Republic
International Clinical Research Center St Anne's University Hospital Brno Czech Republic
Citace poskytuje Crossref.org
Left ventricular septal pacing - can we trust the ECG?
Pacing of the specialized His-Purkinje conduction system: 'back to the future'