BACKGROUND: Inappropriate therapy (IAT) is an undesirable side effect of implantable cardiac defibrillator (ICD) therapy. Early studies with the subcutaneous ICD (S-ICD) showed relatively high inappropriate shock (IAS) rates. The PRAETORIAN (Prospective Randomized Comparison of Subcutaneous and Transvenous Implantable Cardioverter Defibrillator Therapy) trial demonstrated that the S-ICD is noninferior to the transvenous ICD (TV-ICD) with regard to the combined end point of IAS and complications. This secondary analyses evaluates all IAT in the PRAETORIAN trial. METHODS: This international, multicenter trial randomized 849 patients with an indication for ICD therapy between S-ICD (n=426) and TV-ICD therapy (n=423). ICD programming was mandated by protocol. All analysis were performed in the modified intention-to-treat population. RESULTS: In both groups 42 patients experienced IAT (48-month Kaplan-Meier estimated cumulative incidence, 9.9% and 10.1%, respectively; hazard ratio (HR), 0.99 [95% CI, 0.65-1.52]; P=0.97). There was no significant difference in patients experiencing IAS between both groups (P=0.14). In the S-ICD group, 81 IAT episodes with 124 IAS and 1 inappropriate antitachycardia pacing occurred versus 89 IAT episodes with 130 IAS and 124 inappropriate antitachycardia pacing in the TV-ICD group. IAT episodes were most frequently caused by supraventricular tachycardias in the TV-ICD group (n=83/89) versus cardiac oversensing in the S-ICD group (n=40/81). In the TV-ICD group, a baseline heart rate >80 bpm (HR, 1.99 [95% CI, 1.05-3.76]; P=0.03), a history of atrial fibrillation (HR, 2.66 [95% CI, 1.41-5.02]; P=0.003), and smoking (HR, 2.46 [95% CI, 1.31-4.09]; P=0.005) were independent predictors for IAT. A QRS duration >120 ms was an independent predictor for IAT caused by cardiac oversensing in the S-ICD group (HR, 3.13 [95% CI, 1.34-7.31]; P=0.008). Post-IAS interventions significantly reduced IAS recurrence in both groups (P=0.046). CONCLUSIONS: There was no significant difference in IAT and IAS rates between the S-ICD and TV-ICD in a conventional ICD population, but causes and predictors for IAT differed between the devices. After the first IAS, an intervention significantly reduced the recurrence rate of IAS. REGISTRATION: URL: https://www.clinicaltrials.gov; Unique identifier: NCT01296022.
- MeSH
- časové faktory MeSH
- defibrilátory implantabilní * MeSH
- elektrická defibrilace * přístrojové vybavení škodlivé účinky MeSH
- lidé středního věku MeSH
- lidé MeSH
- prospektivní studie MeSH
- rizikové faktory MeSH
- senioři MeSH
- výsledek terapie MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- randomizované kontrolované studie MeSH
- srovnávací studie MeSH
Leadless pacemakers (LPs) are self-contained pacemakers implanted inside the heart, providing a clinical strategy of pacing without pacemaker leads or a subcutaneous pocket. From an experimental therapy first used clinically in 2012, a decade later this technology is an established treatment option. Because of technologic advances and growing evidence, LPs are increasingly being used. Herein, the experience gained from a decade of leadless pacing is reviewed. We cover the safety and efficacy of single-chamber LPs, including comparisons with transvenous pacemakers and various models, and the initial clinical results of the first dual-chamber LP system. Furthermore, evidence and considerations regarding the optimal replacement strategy will be covered. Finally, we discuss future device developments that may broaden indications, such as LPs communicating with subcutaneous implantable cardiac defibrillators and energy-harvesting LPs.
- MeSH
- design vybavení * MeSH
- kardiostimulace umělá * metody trendy MeSH
- kardiostimulátor * trendy MeSH
- lidé MeSH
- srdeční arytmie terapie MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- přehledy MeSH
BACKGROUND: The subcutaneous implantable cardioverter-defibrillator (S-ICD) was developed to overcome the risk of lead-related complications associated with the transvenous implantable cardioverter-defibrillator (TV-ICD). In contrast to the TV-ICD, the S-ICD is a completely extrathoracic device. Subsequently, complications differ between these 2 implantable cardioverter-defibrillators, which might impact patient perceptions of the therapies. This prespecified secondary analysis of the PRAETORIAN trial evaluates differences in quality of life. METHODS: The PRAETORIAN trial (A Prospective, Randomized Comparison of Subcutaneous and Transvenous Implantable Cardioverter Defibrillator Therapy) randomized patients with an implantable cardioverter-defibrillator indication, without the need for pacing to S-ICD or TV-ICD therapy. Two questionnaires were collected at baseline, discharge, 12 months, and 30 months. The Duke Activity Status Index measures cardiac-specific physical functioning, and the 36-Item Short Form Health Survey measures physical and mental well-being, with the subscales bodily pain and mental health being of interest in this analysis. Mann-Whitney U tests were used to compare study arms, and a mixed model was used to describe the questionnaire outcomes over time. RESULTS: Patients were randomized to S-ICD (n=426) and TV-ICD (n=423). In the S-ICD group, 20% were women versus 19% in the TV-ICD group. The median age was 63 (interquartile range, 54-69) years in the S-ICD group versus 64 (interquartile range, 56-69) years in the TV-ICD group. There were no significant differences in the Duke Activity Status Index and 36-Item Short Form Health Survey subscales for bodily pain and mental health between the groups at any time point. Patients with a shock in the last 90 days had significantly lower scores for social functioning (P=0.008) and role limitations due to emotional problems (P=0.001) than patients without a shock, but this effect did not differ between treatment arms. CONCLUSIONS: In a large randomized cohort of patients with an S-ICD or TV-ICD, no difference in overall quality of life was observed. However, implantable cardioverter-defibrillator shocks resulted in a reduction in quality of life, regardless of the device type or appropriateness. REGISTRATION: URL: https://www.clinicaltrials.gov; Unique identifier: NCT01296022.
- MeSH
- časové faktory MeSH
- defibrilátory implantabilní * MeSH
- duševní zdraví * MeSH
- elektrická defibrilace * přístrojové vybavení škodlivé účinky MeSH
- funkční status MeSH
- kvalita života * MeSH
- lidé středního věku MeSH
- lidé MeSH
- náhlá srdeční smrt prevence a kontrola MeSH
- prospektivní studie MeSH
- protézy - design MeSH
- průzkumy a dotazníky MeSH
- rizikové faktory MeSH
- senioři MeSH
- srdeční arytmie terapie diagnóza patofyziologie MeSH
- výsledek terapie MeSH
- zdravotní stav MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- randomizované kontrolované studie MeSH
- srovnávací studie MeSH
In a pivotal trial (NCT05252702), the AVEIR DR (Abbott) leadless pacemaker system was found to be safe and effective in delivering DDDR synchronous atrial and ventricular pacing. This dual-chamber system employs 2 leadless pacemakers with implant-to-implant communication. Although implantation of the ventricular device as a single-chamber pacemaker has been well described, there are additional considerations surrounding the dual-chamber implantation procedure. Herein, we review the dual-chamber leadless pacemaker implantation workflow while providing guidance to optimize safe and effective implantation procedures.
- MeSH
- design vybavení * MeSH
- implantace protézy metody přístrojové vybavení MeSH
- kardiostimulace umělá metody MeSH
- kardiostimulátor * MeSH
- lidé MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- přehledy MeSH
BACKGROUND: The subcutaneous implantable cardioverter-defibrillator (ICD) is associated with fewer lead-related complications than a transvenous ICD; however, the subcutaneous ICD cannot provide bradycardia and antitachycardia pacing. Whether a modular pacing-defibrillator system comprising a leadless pacemaker in wireless communication with a subcutaneous ICD to provide antitachycardia and bradycardia pacing is safe remains unknown. METHODS: We conducted a multinational, single-group study that enrolled patients at risk for sudden death from ventricular arrhythmias and followed them for 6 months after implantation of a modular pacemaker-defibrillator system. The safety end point was freedom from leadless pacemaker-related major complications, evaluated against a performance goal of 86%. The two primary performance end points were successful communication between the pacemaker and the ICD (performance goal, 88%) and a pacing threshold of up to 2.0 V at a 0.4-msec pulse width (performance goal, 80%). RESULTS: We enrolled 293 patients, 162 of whom were in the 6-month end-point cohort and 151 of whom completed the 6-month follow-up period. The mean age of the patients was 60 years, 16.7% were women, and the mean (±SD) left ventricular ejection fraction was 33.1±12.6%. The percentage of patients who were free from leadless pacemaker-related major complications was 97.5%, which exceeded the prespecified performance goal. Wireless-device communication was successful in 98.8% of communication tests, which exceeded the prespecified goal. Of 151 patients, 147 (97.4%) had pacing thresholds of 2.0 V or less, which exceeded the prespecified goal. The percentage of episodes of arrhythmia that were successfully terminated by antitachycardia pacing was 61.3%, and there were no episodes for which antitachycardia pacing was not delivered owing to communication failure. Of 162 patients, 8 died (4.9%); none of the deaths were deemed to be related to arrhythmias or the implantation procedure. CONCLUSIONS: The leadless pacemaker in wireless communication with a subcutaneous ICD exceeded performance goals for freedom from major complications related to the leadless pacemaker, for communication between the leadless pacemaker and subcutaneous ICD, and for the percentage of patients with a pacing threshold up to 2.0 V at a 0.4-msec pulse width at 6 months. (Funded by Boston Scientific; MODULAR ATP ClinicalTrials.gov NCT04798768.).
- MeSH
- bezdrátová technologie MeSH
- bradykardie * terapie MeSH
- defibrilátory implantabilní * škodlivé účinky MeSH
- design vybavení MeSH
- kardiostimulace umělá škodlivé účinky metody MeSH
- kardiostimulátor * škodlivé účinky MeSH
- lidé středního věku MeSH
- lidé MeSH
- náhlá srdeční smrt * prevence a kontrola etiologie MeSH
- následné studie MeSH
- senioři MeSH
- srdeční arytmie * komplikace terapie MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- klinické zkoušky MeSH
- multicentrická studie MeSH
BACKGROUND: The first dual-chamber leadless pacemaker (DC-LP) system consists of 2 separate atrial and ventricular devices that communicate to maintain synchronous atrioventricular pacing and sensing. The initial safety and efficacy were previously reported. OBJECTIVE: The purpose of this study was to evaluate the chronic electrical performance of the DC-LP system. METHODS: Patients meeting standard dual-chamber pacing indications were enrolled and implanted with the DC-LP system (Aveir DR, Abbott), including right atrial and ventricular helix-fixation LPs (atrial leadless pacemaker [ALP], ventricular leadless pacemaker [VLP]). Pacing capture threshold, sensed amplitude, and pacing impedance were collected using the device programmer at prespecified timepoints from 0-6 months postimplant. RESULTS: De novo devices were successfully implanted in 381 patients with complete 6-month data (62% male; age 69 ± 14 years; weight 82 ± 20 kg; 65% sinus nodal dysfunction, 30% atrioventricular block). ALPs were implanted predominantly in the right atrial appendage anterior base and VLPs primarily at the mid-to-apical right ventricular septum. From implant to 1 month, pacing capture thresholds (0.4-ms pulse width) improved in both ALPs (2.4 ± 1.5 V to 0.8 ± 0.8 V; P <.001) and VLPs (0.8 ± 0.6 V to 0.6 ± 0.4 V; P <.001). Sensed amplitudes improved in both ALPs (1.8 ± 1.3 mV to 3.4 ± 1.9 mV; P <.001) and VLPs (8.8 ± 4.0 mV to 11.7 ± 4.2 mV; P <.001). Impedances were stable in ALPs (334 ± 68 Ω to 329 ± 52 Ω; P = .17) and reduced in VLPs (789 ± 351 Ω to 646 ± 190 Ω; P <.001). Electrical measurements remained relatively stable from 1-6 months postimplant. No differences in electrical metrics were observed among ALP or VLP implant locations. CONCLUSION: This first in-human evaluation of the new dual-chamber leadless pacemaker system demonstrated reliable electrical performance throughout the initial 6-month evaluation period.
- MeSH
- atrioventrikulární blokáda terapie patofyziologie MeSH
- časové faktory MeSH
- design vybavení * MeSH
- kardiostimulace umělá * metody MeSH
- kardiostimulátor * MeSH
- lidé MeSH
- následné studie MeSH
- senioři MeSH
- srdeční komory patofyziologie MeSH
- syndrom chorého sinu terapie patofyziologie MeSH
- výsledek terapie MeSH
- Check Tag
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
BACKGROUND: The clinical utilization of leadless pacemakers (LPs) as an alternative to traditional transvenous pacemakers is likely to increase with the advent of dual-chamber LP systems. Since device retrieval to allow LP upgrade or replacement will become an important capability, the first such dual-chamber, helix-fixation LP system (Aveir DR; Abbott, Abbott Park, IL) was specifically designed to allow catheter-based retrieval. In this study, the preclinical performance and safety of retrieving chronically implanted dual-chamber LPs was evaluated. METHODS: Atrial and ventricular LPs were implanted in the right atrial appendage and right ventricular apex of 9 healthy ovine subjects. After ≈2 years, the LPs were retrieved using a dedicated transvenous retrieval catheter (Aveir Retrieval Catheter; Abbott) by snaring, docking, and unscrewing from the myocardium. Comprehensive necropsy/histopathology studies were conducted to evaluate device- and procedure-related outcomes. RESULTS: At a median of 1.9 years postimplant (range, 1.8-2.6), all 18 of 18 (100%) LPs were retrieved from 9 ovine subjects without complications. The median retrieval procedure duration for both LPs, from first-catheter-in to last-catheter-out, was 13.3 minutes (range, 2.5-36.4). Postretrieval, all right atrial, and right ventricular implant sites demonstrated minimal tissue disruption, with intact fibrous tissue limited to the distal device body. No significant device-related trauma, perforation, pericardial effusion, right heart or tricuspid valve injury, or chronic pulmonary thromboembolism were observed at necropsy. CONCLUSIONS: This preclinical study demonstrated the safe and effective retrieval of chronically implanted, helix-fixation, dual-chamber LP systems, paving the way for clinical studies of LP retrieval.
- MeSH
- design vybavení MeSH
- fibrilace síní * MeSH
- kardiostimulátor * MeSH
- lidé MeSH
- lipopolysacharidy MeSH
- ovce domácí MeSH
- ovce MeSH
- srdeční komory MeSH
- zvířata MeSH
- Check Tag
- lidé MeSH
- zvířata MeSH
- Publikační typ
- časopisecké články MeSH
BACKGROUND: The subcutaneous implantable cardioverter-defibrillator (S-ICD) has demonstrated safety and efficacy for the treatment of malignant ventricular arrhythmias. However, a limitation of the S-ICD lies in the inability to either pace-terminate ventricular tachycardia or provide prolonged bradycardia pacing support. OBJECTIVE: The rationale and design of a prospective, single-arm, multinational trial of an intercommunicative leadless pacing system integrated with the S-ICD will be presented. METHODS: A technical description of the modular cardiac rhythm management (mCRM) system (EMPOWER leadless pacemaker and EMBLEM S-ICD) and the implantation procedure is provided. MODULAR ATP (Effectiveness of the EMPOWERTM Modular Pacing System and EMBLEMTM Subcutaneous ICD to Communicate Antitachycardia Pacing) is a multicenter, international trial enrolling up to 300 patients at risk of sudden cardiac death at up to 60 centers trial design. The safety endpoint of freedom from major complications related to the mCRM system or implantation procedure at 6 months and 2 years are significantly higher than 86% and 81%, respectively, and all-cause survival is significantly >85% at 2 years. RESULTS: Efficacy endpoints are that at 6 months mCRM communication success is significantly higher than 88% and the percentage of subjects with low and stable thresholds is significantly higher than 80%. Substudies to evaluate rate-responsive features and performance of the pacing module are also described. CONCLUSION: The MODULAR ATP global clinical trial will prospectively test the safety and efficacy of the first intercommunicating leadless pacing system with the S-ICD. This trial will allow for robust validation of device-device communication, pacing performance, rate responsiveness, and system safety.
- Publikační typ
- časopisecké články MeSH
BACKGROUND: Single-chamber ventricular leadless pacemakers do not support atrial pacing or consistent atrioventricular synchrony. A dual-chamber leadless pacemaker system consisting of two devices implanted percutaneously, one in the right atrium and one in the right ventricle, would make leadless pacemaker therapy a treatment option for a wider range of indications. METHODS: We conducted a prospective, multicenter, single-group study to evaluate the safety and performance of a dual-chamber leadless pacemaker system. Patients with a conventional indication for dual-chamber pacing were eligible for participation. The primary safety end point was freedom from complications (i.e., device- or procedure-related serious adverse events) at 90 days. The first primary performance end point was a combination of adequate atrial capture threshold and sensing amplitude at 3 months. The second primary performance end point was at least 70% atrioventricular synchrony at 3 months while the patient was sitting. RESULTS: Among the 300 patients enrolled, 190 (63.3%) had sinus-node dysfunction and 100 (33.3%) had atrioventricular block as the primary pacing indication. The implantation procedure was successful (i.e., two functioning leadless pacemakers were implanted and had established implant-to-implant communication) in 295 patients (98.3%). A total of 35 device- or procedure-related serious adverse events occurred in 29 patients. The primary safety end point was met in 271 patients (90.3%; 95% confidence interval [CI], 87.0 to 93.7), which exceeded the performance goal of 78% (P<0.001). The first primary performance end point was met in 90.2% of the patients (95% CI, 86.8 to 93.6), which exceeded the performance goal of 82.5% (P<0.001). The mean (±SD) atrial capture threshold was 0.82±0.70 V, and the mean P-wave amplitude was 3.58±1.88 mV. Of the 21 patients (7%) with a P-wave amplitude of less than 1.0 mV, none required device revision for inadequate sensing. At least 70% atrioventricular synchrony was achieved in 97.3% of the patients (95% CI, 95.4 to 99.3), which exceeded the performance goal of 83% (P<0.001). CONCLUSIONS: The dual-chamber leadless pacemaker system met the primary safety end point and provided atrial pacing and reliable atrioventricular synchrony for 3 months after implantation. (Funded by Abbott Medical; Aveir DR i2i ClinicalTrials.gov number, NCT05252702.).
- MeSH
- atrioventrikulární blokáda terapie MeSH
- fibrilace síní terapie etiologie MeSH
- kardiostimulace umělá * škodlivé účinky metody MeSH
- kardiostimulátor * škodlivé účinky MeSH
- lidé MeSH
- následné studie MeSH
- prospektivní studie MeSH
- srdeční arytmie * terapie MeSH
- výsledek terapie MeSH
- Check Tag
- lidé MeSH
- Publikační typ
- časopisecké články MeSH
- klinická studie MeSH
- multicentrická studie MeSH
BACKGROUND: Dual-chamber leadless pacemakers (LPs) require robust communication between distinct right atrial (RA) and right ventricular (RV) LPs to achieve atrioventricular (AV) synchrony. OBJECTIVE: The purpose of this preclinical study was to evaluate a novel, continuous implant-to-implant (i2iTM) communication methodology for maintaining AV-synchronous, dual-chamber DDD(R) pacing by the 2 LPs. METHODS: RA and RV LPs were implanted and paired in 7 ovine subjects (4 with induced complete heart block). AV synchrony (% AV intervals <300 ms) and i2i communication success (% successful i2i transmissions between LPs) were evaluated acutely and chronically. During acute testing, 12-lead electrocardiographic and LP diagnostic data were collected from 5-minute recordings, in 4 postures and 2 rhythms (AP-VP and AS-VP, or AP-VS and AS-VS) per subject. Chronic i2i performance was evaluated through 23 weeks postimplant (final i2i evaluation period: week 16-23). RESULTS: Acute AV synchrony and i2i communication success across multiple postures and rhythms were median [interquartile range] 100.0% [100.0%-100.0%] and 99.9% [99.9%-99.9%], respectively. AV synchrony and i2i success rates did not differ across postures (P = .59, P = .11) or rhythms (P = 1, P = .82). During the final i2i evaluation period, the overall i2i success was 98.9% [98.1%-99.0%]. CONCLUSION: Successful AV-synchronous, dual-chamber DDD(R) leadless pacing using a novel, continuous, wireless communication modality was demonstrated across variations in posture and rhythm in a preclinical model.