Background: The number of implantable cardioverter defibrillator (ICD) infections is increasing due to an increased number of ICD implants, higher-risk patients, and more frequent replacement procedures, which carry a higher risk of infection. Reducing the morbidity, mortality, and cost of ICD-related infections requires an understanding of the current rate of this complication and its predictors. Methods: The Shock Implant Evaluation Trial (SIMPLE) trial randomized 2500 ICD recipients to defibrillation testing or not. Over an average of 3.1 years, patients were seen every 6 months and examined for evidence of ICD infection, which was defined as requiring device removal and/or intravenous antibiotics. Results: Within 24 months, 21 patients (0.8%) developed infection. Fourteen patients (67%) with infection presented within 30 days, 20 patients by 12 months, and only 1 patient beyond 12 months. Univariate analysis demonstrated that patients with primary electrical disorders (3 patients, P = 0.009) and those with a secondary prevention indication (13 patients, P = 0.0009) were more likely to develop infection. Among the 2.2% of patients who developed an ICD wound hematoma, 10.4% developed an infection. Among the 8.3% of patients requiring an ICD reintervention, 1.9% developed an infection. Conclusions: This cohort of ICD recipients at high-volume centres have a low risk of device-related infection. However; strategies to reduce wound hematoma and the need for ICD reintervention could further reduce the rate of infection.
- Publikační typ
- časopisecké články MeSH
BACKGROUND: The Shockless IMPLant Evaluation (SIMPLE) trial showed that defibrillation testing (DT) at the time of implantable cardioverter-defibrillator (ICD) implant did not improve shock efficacy or reduce mortality. There are no data regarding the risk of complications, including stroke, among patients with atrial fibrillation (AF) who undergo DT. OBJECTIVE: The purpose of this prospectively planned substudy of SIMPLE was to evaluate the effect of DT vs no DT on clinical outcomes among patients with AF. METHODS: We compared efficacy (failed appropriate shock/arrhythmic death) and safety between patients who had AF on their immediate preprocedural ECG to the rest of the study patients. Then among patients with AF we compared these outcomes between patients randomized to DT vs no DT. RESULTS: Of the 2500 patients enrolled in SIMPLE, 251 (10%) were in AF immediately before ICD implant. AF patients had an increased risk of failed appropriate shock/arrhythmic death (adjusted hazard ratio [HR] 1.64; 95% confidence interval [CI] 1.13-2.39; P = .009) and higher all-cause mortality (adjusted HR 1.58; 95% CI 1.2-2.08; P = .001). Among AF patients, perioperative complications and stroke did not significantly differ between DT vs no-DT groups (9.2% vs 5.4%; P = .2; and 1.7% vs 1.5%; P >.999, respectively). Failed appropriate shock or arrhythmic death occurred in 35 of 251 AF patients (14%), and the no-DT group proved not inferior to the DT group (HR 0.58; 95% CI 0.30-1.15; Pnoninferiority = .006). CONCLUSION: ICD recipients with AF are at increased risk for adverse outcomes; however, DT does not improve arrhythmic survival or shock efficacy. There is no evidence that DT increased the occurrence of perioperative stroke.
- MeSH
- defibrilátory implantabilní * MeSH
- elektrokardiografie * MeSH
- fibrilace síní mortalita patofyziologie terapie MeSH
- lidé středního věku MeSH
- lidé MeSH
- míra přežití trendy MeSH
- následné studie MeSH
- převodní systém srdeční patofyziologie MeSH
- prospektivní studie MeSH
- senioři MeSH
- výsledek terapie MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- randomizované kontrolované studie MeSH
- Geografické názvy
- Evropa MeSH
INTRODUCTION: Dual-coil leads (DC-leads) were the standard of choice since the first nonthoracotomy implantable cardioverter/defibrillator (ICD). We used contemporary data to determine if DC-leads offer any advantage over single-coil leads (SC-leads), in terms of defibrillation efficacy, safety, clinical outcome, and complication rates. METHODS AND RESULTS: In the Shockless IMPLant Evaluation study, 2500 patients received a first implanted ICD and were randomized to implantation with or without defibrillation testing. Two thousand and four hundred seventy-five patients received SC-coil or DC-coil leads (SC-leads in 1025/2475 patients; 41.4%). In patients who underwent defibrillation testing (n = 1204), patients with both lead types were equally likely to achieve an adequate defibrillation safety margin (88.8% vs 91.2%; P = 0.16). There was no overall effect of lead type on the primary study endpoint of "failed appropriate shock or arrhythmic death" (adjusted HR 1.18; 95% CI, 0.86-1.62; P = 0.300), and on all-cause mortality (SC-leads: 5.34%/year; DC-leads: 5.48%/year; adjusted HR 1.16; 95% CI, 0.94-1.43; P = 0.168). However, among patients without prior heart failure (HF), and SC-leads had a significantly higher risk of failed appropriate shock or arrhythmic death (adjusted HR 7.02; 95% CI, 2.41-20.5). There were no differences in complication rates. CONCLUSION: In this nonrandomized evaluation, there was no overall difference in defibrillation efficacy, safety, outcome, and complication rates between SC-leads and DC-leads. However, DC-leads were associated with a reduction in the composite of failed appropriate shock or arrhythmic death in the subgroup of non-HF patients. Considering riskier future lead extraction with DC-leads, SC-leads appears to be preferable in the majority of patients.
- MeSH
- časové faktory MeSH
- defibrilátory implantabilní * MeSH
- elektrická defibrilace škodlivé účinky přístrojové vybavení mortalita MeSH
- lidé středního věku MeSH
- lidé MeSH
- náhlá srdeční smrt prevence a kontrola MeSH
- příčina smrti MeSH
- primární prevence přístrojové vybavení MeSH
- protézy - design MeSH
- randomizované kontrolované studie jako téma MeSH
- rizikové faktory MeSH
- sekundární prevence přístrojové vybavení MeSH
- selhání protézy MeSH
- senioři MeSH
- srdeční arytmie diagnóza mortalita patofyziologie terapie MeSH
- výsledek terapie MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- pozorovací studie MeSH
- srovnávací studie MeSH
BACKGROUND: Patients with hypertrophic cardiomyopathy (HCM) are considered to be at high risk for elevated defibrillation thresholds, periprocedural complications, and failed appropriate shocks. OBJECTIVE: The purpose of this study was to determine the value of defibrillation testing (DT) in HCM patients undergoing implantable cardioverter-defibrillator (ICD) insertion. METHODS: Defibrillation thresholds, perioperative complications, and long-term outcomes were compared between patients with HCM and those with ischemic cardiomyopathy (ICM) or dilated cardiomyopathy (DCM) enrolled in the SIMPLE (Shockless IMPLant Evaluation) trial (Clinialtrials.gov Identifier: NCT00800384). In patients with HCM, outcomes were also compared between those randomized to DT vs no DT. RESULTS: Adequate defibrillation safety margin without system change was achieved in 46 of 52 (88.5%) HCM and 948 of 1047 (90.5%) ICM/DCM patients (P = .63). Perioperative complications occurred in 1 of 52 (1.9%) HCM patients with DT compared to 67 of 1047 (6.4%) ICM/DCM patients with DT (P = .37) or 3 of 42 (7.1%) HCM patients without DT (P = .32). During follow-up, there was no significant difference between HCM vs ICM/DCM patients in terms of all-cause mortality (adjusted hazard ratio [HR] 1.02, 95% confidence interval [CI] 0.45-2.34), composite of arrhythmic death or failed appropriate shock (adjusted HR 0.33, 95% CI 0.04-2.42), inappropriate shocks (adjusted HR 1.64, 95% CI 0.69-3.89), or system complications (adjusted HR 1.93, 95% CI 0.88-4.27). All-cause mortality (HR 0.26, 95% CI 0.03-2.20), appropriate (HR 0.24, 95% CI 0.03-2.05), and inappropriate shocks (HR 2.13, 95% CI 0.51-8.94) were similar in HCM patients without or those with DT. CONCLUSION: We did not find any difference in intraoperative defibrillation efficacy, perioperative complications, and long-term outcomes between patients with HCM and those with ICM/DCM. DT did not improve intraoperative or clinical shock efficacy in HCM patients.
- MeSH
- celosvětové zdraví MeSH
- defibrilátory implantabilní * MeSH
- elektrokardiografie MeSH
- hypertrofická kardiomyopatie mortalita patofyziologie terapie MeSH
- incidence MeSH
- jednoduchá slepá metoda MeSH
- lidé středního věku MeSH
- lidé MeSH
- míra přežití trendy MeSH
- náhlá srdeční smrt epidemiologie prevence a kontrola MeSH
- prognóza MeSH
- výsledek terapie MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- práce podpořená grantem MeSH
- randomizované kontrolované studie MeSH
Aims: No standard practice exists with respect to anaesthesiologist-directed sedation (ADS) vs. sedation by proceduralist (PDS) for defibrillation threshold (DT) testing. We aimed to evaluate adverse events and safety outcomes with ADS vs. PDS for DT testing. Methods and results: A post hoc analysis of the Shockless Implant Evaluation (SIMPLE) study was performed among the 1242 patients who had DT testing (624 ADS and 618 PDS). We evaluated both intraoperative and in-hospital adverse composite events and two safety composite outcomes at 30-days of the main trial. Propensity score adjusted models were used to compute odds ratio (OR) and 95% confidence interval (CI) to evaluate the association between adverse and safety outcomes with method of sedation and independent predictors for use of ADS. Compared to PDS, patients who received ADS were younger (62 ± 12 years vs. 64 ± 12 years, P = 0.01), had lower ejection fraction (left ventricular ejection fraction 0.31 ± 13 vs. 0.33 ± 13, P = 0.03), were more likely to receive inhalational anaesthesia, propofol, or narcotics (P < 0.001, respectively) and receive an arterial line (43% vs. 8%, P = <0.0001). Independent predictors for ADS sedation were presence of coronary artery disease (OR 1.69, 95% CI 1.0-2.72; P = 0.03) and hypertrophic cardiomyopathy (OR 2.64, 95% CI 1.19-5.85; P = 0.02). Anaesthesiologist directed sedation had higher intraoperative adverse events (2.2% vs. 0.5%; OR 4.47, 95% CI 1.25-16.0; P = 0.02) and higher primary safety outcomes at 30 days (8.2% vs. 4.9%; OR 1.72 95% CI 1.06-2.80; P = 0.03) and no difference in other outcomes compared to PDS. Conclusion: Proceduralist-directed sedation is safe, however, this could be result of selection bias. Further research is needed.
- MeSH
- analgosedace * škodlivé účinky metody MeSH
- defibrilátory implantabilní * MeSH
- fibrilace komor terapie MeSH
- hodnocení rizik MeSH
- hypnotika a sedativa * aplikace a dávkování škodlivé účinky MeSH
- implantace protézy * škodlivé účinky přístrojové vybavení metody MeSH
- lidé středního věku MeSH
- lidé MeSH
- náhlá srdeční smrt etiologie prevence a kontrola MeSH
- pooperační komplikace * diagnóza epidemiologie etiologie MeSH
- rizikové faktory MeSH
- senioři MeSH
- výsledky a postupy - zhodnocení (zdravotní péče) MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- randomizované kontrolované studie MeSH
Aims: Pocket haematoma is a common complication after defibrillator [implantable cardioverter defibrillator (ICD)] implantation, which is not only painful, but also increases the risk of device-related infection, and possibly embolic events. The present study seeks to evaluate the rate and predictors of clinically significant pocket haematoma. Methods and results: This study included 2500 patients receiving an ICD in the SIMPLE trial. A clinically significant pocket haematoma was defined as a haematoma that required re-operation or interruption of oral anticoagulation (OAC) therapy. Clinically significant pocket haematoma occurred in 56 of 2500 patients (2.2%) of which 6 (10.7%) developed device-related infection. Patients who developed pocket haematoma were older (mean age 67.6 ± 8.8 years vs. 62.7 ± 11.6 years, P < 0.001), were more likely to have permanent atrial fibrillation (30.4 vs. 6.7%, P < 0.001) and a history of stroke (17.9 vs. 6.7%, P = 0.004), or were more likely to receive peri-operative OAC (50.0 vs. 28.4%, P < 0.001), unfractionated heparin (16.1 vs. 5.2%, P = 0.003), or low-molecular-weight heparin (37.5 vs. 17.5%, P < 0.001). Independent predictors of wound haematoma on multivariable analysis included the use of heparin bridging (OR 2.65, 95% CI 1.48-4.73, P = 0.001), sub-pectoral location of ICD (OR 2.00, 95% CI 1.12-3.57, P =0.020), previous stroke (OR 2.47, 95% CI 1.20-5.10, P = 0.015), an upgrade from permanent pacemaker (OR 2.52, 95% CI 1.07-5.94, P = 0.035), and older age (OR 1.03, 95% CI 1.00-1.06, P = 0.049). Conclusion: Pocket haematoma remains an important complication of ICD implantation and is associated with a high risk of infection. Independent predictors of pocket haematoma include heparin bridging, prior stroke, sub-pectoral placement of ICD, older age, and upgrade from a pacemaker.
- MeSH
- chirurgická rána diagnóza epidemiologie MeSH
- defibrilátory implantabilní * škodlivé účinky MeSH
- elektrická defibrilace škodlivé účinky přístrojové vybavení mortalita MeSH
- hematom diagnóza epidemiologie MeSH
- heparin aplikace a dávkování škodlivé účinky MeSH
- implantace protézy škodlivé účinky přístrojové vybavení mortalita MeSH
- incidence MeSH
- infekce spojené s protézou epidemiologie MeSH
- lidé středního věku MeSH
- lidé MeSH
- logistické modely MeSH
- multivariační analýza MeSH
- nelineární dynamika MeSH
- odds ratio MeSH
- prospektivní studie MeSH
- rizikové faktory MeSH
- rozdělení chí kvadrát MeSH
- senioři MeSH
- srdeční arytmie diagnóza mortalita terapie MeSH
- věkové faktory MeSH
- výsledek terapie MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- randomizované kontrolované studie MeSH
INTRODUCTION: We aim to compare the acute and long-term success of defibrillation between non-apical and apical ICD lead position. METHODS AND RESULTS: The position of the ventricular lead was recorded by the implanting physician for 2,475 of 2,500 subjects in the Shockless IMPLant Evaluation (SIMPLE) trial, and subjects were grouped accordingly as non-apical or apical. The success of intra-operative defibrillation testing and of subsequent clinical shocks were compared. Propensity scoring was used to adjust for the impact of differences in baseline variables between these groups. There were 541 leads that were implanted at a non-apical position (21.9%). Patients implanted with a non-apical lead had a higher rate of secondary prevention indication. Non-apical location resulted in a lower mean R-wave amplitude (14.0 vs. 15.2, P < 0.001), lower mean pacing impedance (662 ohm vs. 728 ohm, P < 0.001), and higher mean pacing threshold (0.70 V vs. 0.66 V, P = 0.01). Single-coil leads and cardiac resynchronization devices were used more often in non-apical implants. The success of intra-operative defibrillation was similar between propensity score matched groups (89%). Over a mean follow-up of 3 years, there were no significant differences in the yearly rates of appropriate shock (5.5% vs. 5.4%, P = 0.98), failed appropriate first shock (0.9% vs. 1.0%, P = 0.66), or the composite of failed shock or arrhythmic death (2.8% vs. 2.3% P = 0.35) according to lead location. CONCLUSION: We did not detect any reduction in the ICD efficacy at the time of implant or during follow-up in patients receiving a non-apical RV lead.
- MeSH
- časové faktory MeSH
- defibrilátory implantabilní * MeSH
- elektrická defibrilace škodlivé účinky přístrojové vybavení metody mortalita MeSH
- elektrofyziologické techniky kardiologické MeSH
- Kaplanův-Meierův odhad MeSH
- kardiostimulace umělá MeSH
- lidé středního věku MeSH
- lidé MeSH
- logistické modely MeSH
- náhlá srdeční smrt etiologie MeSH
- proporcionální rizikové modely MeSH
- prospektivní studie MeSH
- protézy - design MeSH
- rizikové faktory MeSH
- selhání protézy MeSH
- senioři MeSH
- srdeční arytmie komplikace diagnóza mortalita terapie MeSH
- tendenční skóre MeSH
- výsledek terapie MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- senioři MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- práce podpořená grantem MeSH
- randomizované kontrolované studie MeSH
- srovnávací studie MeSH
BACKGROUND: Defibrillation testing by induction and termination of ventricular fibrillation is widely done at the time of implantation of implantable cardioverter defibrillators (ICDs). We aimed to compare the efficacy and safety of ICD implantation without defibrillation testing versus the standard of ICD implantation with defibrillation testing. METHODS: In this single-blind, randomised, multicentre, non-inferiority trial (Shockless IMPLant Evaluation [SIMPLE]), we recruited patients aged older than 18 years receiving their first ICD for standard indications at 85 hospitals in 18 countries worldwide. Exclusion criteria included pregnancy, awaiting transplantation, particpation in another randomised trial, unavailability for follow-up, or if it was expected that the ICD would have to be implanted on the right-hand side of the chest. Patients undergoing initial implantation of a Boston Scientific ICD were randomly assigned (1:1) using a computer-generated sequence to have either defibrillation testing (testing group) or not (no-testing group). We used random block sizes to conceal treatment allocation from the patients, and randomisation was stratified by clinical centre. Our primary efficacy analysis tested the intention-to-treat population for non-inferiority of no-testing versus testing by use of a composite outcome of arrhythmic death or failed appropriate shock (ie, a shock that did not terminate a spontaneous episode of ventricular tachycardia or fibrillation). The non-inferiority margin was a hazard ratio (HR) of 1·5 calculated from a proportional hazards model with no-testing versus testing as the only covariate; if the upper bound of the 95% CI was less than 1·5, we concluded that ICD insertion without testing was non-inferior to ICD with testing. We examined safety with two, 30 day, adverse event outcome clusters. The trial is registered with ClinicalTrials.gov, number NCT00800384. FINDINGS: Between Jan 13, 2009, and April 4, 2011, of 2500 eligible patients, 1253 were randomly assigned to defibrillation testing and 1247 to no-testing, and followed up for a mean of 3·1 years (SD 1·0). The primary outcome of arrhythmic death or failed appropriate shock occurred in fewer patients (90 [7% per year]) in the no-testing group than patients who did receive it (104 [8% per year]; HR 0·86, 95% CI 0·65-1·14; pnon-inferiority <0·0001). The first safety composite outcome occurred in 69 (5·6%) of 1236 patients with no-testing and in 81 (6·5%) of 1242 patients with defibrillation testing, p=0·33. The second, pre-specified safety composite outcome, which included only events most likely to be directly caused by testing, occurred in 3·2% of patients with no-testing and in 4·5% with defibrillation testing, p=0·08. Heart failure needing intravenous treatment with inotropes or diuretics was the most common adverse event (in 20 [2%] of 1236 patients in the no-testing group vs 28 [2%] of 1242 patients in the testing group, p=0·25). INTERPRETATION: Routine defibrillation testing at the time of ICD implantation is generally well tolerated, but does not improve shock efficacy or reduce arrhythmic death. FUNDING: Boston Scientific and the Heart and Stroke Foundation (Ontario Provincial office).
- MeSH
- bezpečnost pacientů MeSH
- defibrilátory implantabilní * MeSH
- elektrická defibrilace metody mortalita MeSH
- fibrilace komor etiologie MeSH
- hodnocení rizik MeSH
- implantace protézy metody mortalita MeSH
- jednoduchá slepá metoda MeSH
- lidé středního věku MeSH
- lidé MeSH
- pooperační komplikace etiologie mortalita MeSH
- prognóza MeSH
- srdeční arytmie mortalita terapie MeSH
- srdeční selhání etiologie MeSH
- Check Tag
- lidé středního věku MeSH
- lidé MeSH
- mužské pohlaví MeSH
- ženské pohlaví MeSH
- Publikační typ
- časopisecké články MeSH
- multicentrická studie MeSH
- práce podpořená grantem MeSH
- randomizované kontrolované studie MeSH
- srovnávací studie MeSH